Salta al contenuto principale
Passa alla visualizzazione normale.

LEONARDO ABBENE

Potentialities of High-Resolution 3-D CZT Drift Strip Detectors for Prompt Gamma-Ray Measurements in BNCT

  • Autori: Abbene L.; Principato F.; Buttacavoli A.; Gerardi G.; Bettelli M.; Zappettini A.; Altieri S.; Auricchio N.; Caroli E.; Zanettini S.; Protti N.
  • Anno di pubblicazione: 2022
  • Tipologia: Articolo in rivista
  • OA Link: http://hdl.handle.net/10447/536038

Abstract

Recently, new high-resolution cadmium–zinc–telluride (CZT) drift strip detectors for room temperature gamma-ray spectroscopic imaging were developed by our group. The CZT detectors equipped with orthogonal anode/cathode collecting strips, drift strips and dedicated pulse processing allow a detection area of 6 × 20 mm2 and excellent room temperature spectroscopic performance (0.82% FWHM at 661.7 keV). In this work, we investigated the potentialities of these detectors for prompt gamma-ray spectroscopy (PGS) in boron neutron capture therapy (BNCT). The detectors, exploiting the measurement of the 478 keV prompt gamma rays emitted by 94%7Li nuclides from the10B(n, α)7Li reaction, are very appealing for the development of single-photon emission computed tomography (SPECT) systems and Compton cameras in BNCT. High-resolution gamma-ray spectra from10B samples under thermal neutrons were measured at the T.R.I.G.A. Mark II research nuclear reactor of the University of Pavia (Italy).