Salta al contenuto principale
Passa alla visualizzazione normale.

VINCENZO TAORMINA

Comparative Study of Human and Automated Screening for Antinuclear Antibodies by Immunofluorescence on HEp-2 Cells

  • Autori: Gorgi Y; Dhaouadi T; Sfar I; Haouami Y; Ben Abdallah T; Raso G; Cascio D; Cipolla M; Taormina V; Fauci A; Brusca I; Friscia G; Benammar Elgaaïed A; Marrakchi Triki R; Gati A; Ben Ahmed M; Louzir H
  • Anno di pubblicazione: 2015
  • Tipologia: Articolo in rivista (Articolo in rivista)
  • OA Link: http://hdl.handle.net/10447/165458

Abstract

Background: Several automated systems had been developed in order to reduce inter-observer variability in indirect immunofluorescence (IIF) interpretation. We aimed to evaluate the performance of a processing system in antinuclear antibodies (ANA) screening on HEp-2 cells. Patients and Methods: This study included 64 ANA-positive sera and 107 ANA-negative sera that underwent IIF on two commercial kits of HEp-2 cells (BioSystems® and Euroimmun®). IIF results were compared with a novel automated interpretation system, the “CyclopusCADImmuno®” (CAD). Results: All ANA-positive sera images were recognized as positive by CAD (sensitivity = 100%), while 17 (15.9%) of the ANA-negative sera images were interpreted as positive (specificity = 84.1%), =0.799 (SD=0.045). Comparison of IIF pattern determination between human and CAD system revealed on HEp-2 (BioSystems®), a complete concordance in 6 (9.37%) sera, a partial concordance (sharing of at least 1 pattern) in 42 (65.6%) cases and in 16 (25%) sera the pattern interpretation was discordant. Similarly, on HEp-2 (Euroimmun®) the concordance in pattern interpretation was total in 5 (7.8%) sera, partial in 39 (60.9%) and absent in 20 (31.25%). For both tested HEp-2 cells kits agreement was enhanced for the most common patterns, homogenous, fine speckled and coarse speckled. While there was an issue in identification of nucleolar, dots and nuclear membranous patterns by CAD. Conclusion: Assessment of ANA by IIF on HEp-2 cells using the automated interpretation system, the “CyclopusCADImmuno®” is a reliable method for positive/negative differentiation. Continuous integration of IIF images would improve the pattern identification by the CAD.