Skip to main content
Passa alla visualizzazione normale.

GIUSEPPE RASO

Portable CdTe detection system for mammographic X-ray spectroscopy

  • Authors: LA MANNA A; L ABBENE; FFAUCI; GERARDI G; G RASO; S DEL SORDO; L BARONE TONGHI; S STUMBO
  • Publication year: 2006
  • Type: Articolo in rivista (Articolo in rivista)
  • Key words: SPECTRA
  • OA Link: http://hdl.handle.net/10447/12673

Abstract

This paper describes a portable apparatus to be utilized in mammographic X-ray spectroscopy under clinical conditions. The system, based on a CdTe solid-state detector, is able to directly measure mammographic X-ray tube spectra. Good system response to monoenergetic photons was measured using X-ray and γ-ray calibration sources (109Cd and 241Am). The measured molybdenum X-ray spectra, in agreement with simulated spectra, show the good spectral capability of the system also at high photon fluence rates, as typical of clinical mammography. Low tailing, no secondary X-ray escape and low pile-up distortions in the measured spectra indicate that this portable system is suitable for mammographic X-ray spectroscopy. © Società Italiana di Fisica.