Salta al contenuto principale
Passa alla visualizzazione normale.

ANTONIO BARTOLOTTA

Combined TL and 10B-alanine ESR dosimetry for BNCT

  • Autori: BARTOLOTTA A; D'OCA MC; LO GIUDICE B; BRAI M; BORIO R; FORINI N; SALVADORI P; MANERA S
  • Anno di pubblicazione: 2004
  • Tipologia: Articolo in rivista (Articolo in rivista)
  • OA Link: http://hdl.handle.net/10447/6349

Abstract

The dosimetric technique described in this paper is based on electron spin resonance (ESR) detectors using an alanine-boric compound acid enriched with 10B, and beryllium oxide thermoluminescent (TL) detectors; with this combined dosimetry, it is possible to discriminate the doses due to thermal neutrons and gamma radiation in a mixed field. Irradiations were carried out inside the thermal column of a TRIGA MARK II water-pool-type research nuclear reactor, also used for Boron Neutron Capture therapy (BNCT) applications, with thermal neutron fluence from 109 to 1014 nth cm-2. The ESR dosemeters using the alanine-boron compound indicated ESR signals about 30-fold stronger than those using only alanine. Moreover, a negligible correction for the gamma contribution, measured with TL detectors, almost insensitive to thermal neutrons, was necessary. Therefore, a simultaneous analysis of our TL and ESR detectors allows discrimination between thermal neutron and gamma doses, as required in BNCT.